FDA Adverse Event Injury Summary report: N

0009613348-2024-013318

MDR report key: 20388145 · Received October 7, 2024

Report

Report Number
0009613348-2024-013318
Event Type
Injury
Date Received
October 7, 2024
Date of Event
July 3, 2024
Report Date
October 7, 2024
Manufacturer
INSTITUT STRAUMANN AG
Product Code
DZE
PMA / PMN Number
K150938
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BATCH NUMBER COULD BE VERIFIED. OUR MANUFACTURING Q-SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFIRM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. THE NON-INTEGRATION OF AN ENDOSSEOUS DENTAL IMPLANT DURING THE HEALING PHASE IS A KNOWN INHERENT RISK OF THE TREATMENT WITH DENTAL IMPLANTS. MOST IMPLANT FAILURES OCCUR BEFORE OCCLUSAL LOADING (ANALAINEN ET AL. 2009). BASED ON CLINICAL STUDIES ABOUT 1-3% OF THE IMPLANTS FAIL WITHIN THE FIRST YEAR AFTER IMPLANTATION (GANELES ET AL. 2008). IMPLANTS MAY HAVE TO BE REMOVED IN CASE ONE OR MORE OF THE IMPLANT SUCCESS CRITERIA ARE NOT MET. IMPLANT SUCCESS CRITERIA ACCORDING TO BUSER ET AL. (1991) ARE: 1. ABSENCE OF PERSISTENT SUBJECTIVE COMPLAINTS SUCH AS PAIN, FOREIGN BODY SENSATION AND /OR DYSESTHESIA 2. ABSENCE OF A RECURRENT PERI-IMPLANT INFECTION WITH SUPPURATION 3. ABSENCE OF IMPLANT MOBILITY 4. ABSENCE OF A CONTINUOUS RADIOLUCENCY AROUND THE IMPLANT. THE MANUFACTURER'S TREND ANALYSIS CONFIRMS THAT THE REPORTED EARLY FAILURE RATE ASSOCIATED WITH ITS DENTAL IMPLANTS IS BELOW THE EXPECTED FAILURE RATE FOR THIS TREATMENT AS PUBLISHED IN THE SCIENTIFIC LITERATURE.

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024 IN ADA 19. DETAILS OF SURGERY: PRIMARY STABILITY NOT ACHIEVED AND IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2024 NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: BLEEDING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
246494 DZE INSTITUT STRAUMANN AG

Patients

Seq Age Sex Outcome Treatment
1 28 YR Male Required Intervention