FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 20383365 · Received October 7, 2024

Report

Report Number
1911916-2024-00716
Event Type
Malfunction
Date Received
October 7, 2024
Date of Event
September 19, 2024
Report Date
October 11, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PR (B)(4) FOLLOW UP FOR DEVICE EVALUATION. IT WAS REPORTED THE SYRINGE DID NOT HAVE MARKINGS ON IT. TO AID IN THE INVESTIGATION, ONE SAMPLE IN AN OPENED PACKAGING BLISTER AND TWO PHOTOS WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. A VISUAL INSPECTION WAS PERFORMED, AND THE SYRINGE BARREL IS MISSING THE SCALE MARKING. THE TWO PHOTOS PROVIDED SHOW THE SAMPLE RECEIVED. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR IF THERE WAS A JAM MISALIGNING THE PRINTING PAD DURING THE SYRINGE BARREL PRINTING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT 4080219. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. THE SAMPLE WILL BE SHOWN TO ASSOCIATES FOR AWARENESS. BASED ON THE INVESTIGATION AND WITH THE RETURNED SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED. MATERIAL # 309653; BATCH # 4080219. IT WAS REPORTED BY CUSTOMER THAT THE 50ML SYRINGE NOT HAVING MARKINGS ON IT. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. CHILDREN'S MINNESOTA HAS AGAIN EXPERIENCED ISSUES WITH THE 50ML SYRINGE NOT HAVING MARKINGS ON IT. PLEASE SEE PRODUCT INFO BELOW AND PHOTO ATTACHED. I DO HAVE THE PRODUCT SAMPLE IF NEEDED FOR INVESTIGATION. BD 50 ML LUER LOK SYRINGE: MFG # 309653. LOT # 4080219.

Description of Event or Problem · 0

MATERIAL # 309653, BATCH # 4080219. IT WAS REPORTED BY CUSTOMER THAT THE 50ML SYRINGE NOT HAVING MARKINGS ON IT. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. (B)(6) HAS AGAIN EXPERIENCED ISSUES WITH THE 50ML SYRINGE NOT HAVING MARKINGS ON IT. PLEASE SEE PRODUCT INFO BELOW AND PHOTO ATTACHED. I DO HAVE THE PRODUCT SAMPLE IF NEEDED FOR INVESTIGATION. BD 50 ML LUER LOK SYRINGE. MFG # 309653. LOT # 4080219.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2541797 SYRINGE 50ML LL TIP 1ML SYRINGE, PISTON FMF BECTON DICKINSON 4080219 30382903096535

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown