FDA Adverse Event Injury Summary report: N

3008261720-2024-003038

MDR report key: 20374894 · Received October 4, 2024

Report

Report Number
3008261720-2024-003038
Event Type
Injury
Date Received
October 4, 2024
Date of Event
November 5, 2023
Report Date
October 4, 2024
Manufacturer
JJGC INDUSTRIA E COMERCIO DE MATERIAIS DENTARIOS SA
Product Code
DZE
PMA / PMN Number
K163194
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BATCH NUMBER COULD NOT BE VERIFIED DUE TO INCOMPLETE OR MISSING INFORMATION AND / OR PRODUCT FROM THE CUSTOMER. OUR MANUFACTURING Q-SYSTEM ASSURES THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFIRM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. THE NON-INTEGRATION OF AN ENDOSSEOUS DENTAL IMPLANT DURING THE HEALING PHASE IS A KNOWN INHERENT RISK OF THE TREATMENT WITH DENTAL IMPLANTS. MOST IMPLANT FAILURES OCCUR BEFORE OCCLUSAL LOADING (ANALAINEN ET AL. 2013, KOLDSLAND ET AL. 2009). BASED ON CLINICAL STUDIES ABOUT 1-3% OF THE IMPLANTS FAIL WITHIN THE FIRST YEAR AFTER IMPLANTATION (GANELES ET AL. 2008). IMPLANTS MAY HAVE TO BE REMOVED IN CASE ONE OR MORE OF THE IMPLANT SUCCESS CRITERIA ARE NOT MET. IMPLANT SUCCESS CRITERIA ACCORDING TO BUSER ET AL. (1991) ARE: 1. ABSENCE OF PERSISTENT SUBJECTIVE COMPLAINTS SUCH AS PAIN, FOREIGN BODY SENSATION AND /OR DYSESTHESIA 2. ABSENCE OF A RECURRENT PERI-IMPLANT INFECTION WITH SUPPURATION 3. ABSENCE OF IMPLANT MOBILITY 4. ABSENCE OF A CONTINUOUS RADIOLUCENCY AROUND THE IMPLANT. THE MANUFACTURER'S TREND ANALYSIS CONFIRMS THAT THE REPORTED EARLY FAILURE RATE ASSOCIATED WITH ITS DENTAL IMPLANTS IS BELOW THE EXPECTED FAILURE RATE FOR THIS TREATMENT AS PUBLISHED IN THE SCIENTIFIC LITERATURE. NON-FATAL SERIOUS INJURY OR DEVICE MALFUNCTION STORED DUE TO COVID 19 PANDEMIC IN ACCORDANCE WITH FDA GUIDANCE "POSTMARKETING ADVERSE EVENT REPORTING FOR MEDICAL PRODUCTS AND DIETARY SUPPLEMENTS DURING A PANDEMIC" PUBLISHED MAY 2020.

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN THE PATIENT'S MOUTH. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2602472 DZE JJGC INDUSTRIA E COMERCIO DE MATERIAIS DENTARIOS SA

Patients

Seq Age Sex Outcome Treatment
1 NA Male Required Intervention