MITRACLIP
Report
- Report Number
- 2135147-2024-04903
- Event Type
- Injury
- Date Received
- October 4, 2024
- Date of Event
- September 12, 2024
- Report Date
- November 1, 2024
- Manufacturer
- ABBOTT MEDICAL
- Product Code
- DRA
- UDI-DI
- 08717648288654
- PMA / PMN Number
- K190167
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ALL AVAILABLE INFORMATION WAS INVESTIGATED, AND THE REPORTED UNINTENDED MOVEMENT ASSOCIATED WITH KNOB SLIPPAGE WAS CONFIRMED VIA DEVICE ANALYSIS. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY IDENTIFIED NO SIMILAR INCIDENTS REPORTED FROM THIS LOT. THE INVESTIGATION DETERMINED THE REPORTED KNOB SLIPPAGE APPEARS TO BE RELATED TO A POTENTIAL PRODUCT QUALITY ISSUE. THE ISSUE APPEARS TO BE RELATED TO AN OPEN TREND EXCEPTION (REFERENCE EXCEPTION (ISSUE) 133059). THE ISSUE IS BEING ADDRESSED PER INTERNAL OPERATING PROCEDURES. ABBOTT WILL CONTINUE TO TREND THE PERFORMANCE OF THESE DEVICES.
THE DEVICE HAS BEEN RECEIVED. HOWEVER, INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THIS WAS A MITRACLIP PROCEDURE TO TREAT FUNCTIONAL MITRAL REGURGITATION WITH A GRADE OF 4. IT WAS NOTED, WHILE APPLYING THE + KNOB, THE KNOB WOULD RETURN TO A NEUTRAL POSITION OVER TIME. THE PHYSICIAN DECIDED TO USE A CLOTH TAPE TO HOLD THE KNOB IN POSITION AND CONTINUE WITH THE PROCEDURE. ONE CLIP WAS THEN IMPLANTED, REDUCING MR TO A GRADE OF 1. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 704422 | MITRACLIP | CATHETER, STEERABLE | DRA | ABBOTT MEDICAL | 40513R1116 | 08717648288654 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Male | Required Intervention | CLIP DELIVERY SYSTEM |