FDA Adverse Event Injury Summary report: N

PERCLOSE¿ PROSTYLE¿

MDR report key: 20371046 · Received October 4, 2024

Report

Report Number
2024168-2024-11724
Event Type
Injury
Date Received
October 4, 2024
Date of Event
September 16, 2024
Report Date
January 8, 2025
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648344039
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PERCLOSE PROSTYLE DEVICES REFERENCED IN B5 ARE FILED UNDER SEPARATE MEDWATCH REPORT NUMBERS. B3: DATE ESTIMATED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. THE FAILURE TO CYCLE WAS CONFIRMED AS A MALPOSITION OF THE DEVICE. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE INVESTIGATION DETERMINED THAT THE DISCOVERED MATERIAL SEPARATION AND REPORTED UNEXPECTED MEDICAL INTERVENTION APPEAR TO BE RELATED TO OPERATIONAL CONTEXT. IT IS LIKELY AN INTERACTION WITH PATIENT ANATOMY OR UNDER INSERTION OF THE DEVICE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4 CORRECTION: LOT # UPDATED FROM 4071442 TO 4062442. H6 CORRECTION: REPORTED DEVICE CODE 3190 REMOVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF THE CALCIFIED COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE PRIOR TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) INTERVENTIONAL PROCEDURE. REPORTEDLY, AN UNKNOWN SUTURE RETRIEVAL ISSUE OCCURRED WITH MULTIPLE PROSTYLE DEVICES. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 14F SHEATH AND THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

SUBSEQUENT TO THE INITIALLY FILED REPORT, THE FOLLOWING INFORMATION WAS RECEIVED. A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED WITH ALL PROSTYLE DEVICES. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1326059 PERCLOSE¿ PROSTYLE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12773-02 4062442 08717648344039

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention