CELL-DYN RUBY SYSTEM
Report
- Report Number
- 2919069-2024-00057
- Event Type
- Malfunction
- Date Received
- October 4, 2024
- Date of Event
- September 12, 2024
- Report Date
- October 28, 2024
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- GKZ
- PMA / PMN Number
- K061667
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CELL-DYN RUBY, SERIAL NUMBER (B)(6) WAS EVALUATED. THE ABBOTT POWER CORD (CORD, PWR, 125VAC/10A), WHICH PLUGGED INTO THE CUSTOMER¿S UNINTERRUPTIBLE POWER SUPPLY (UPS), WAS BURNED. IT WAS SUSPECTED THAT THE ABBOTT POWER CORD (CORD, PWR, 125VAC/10A) MAY NOT HAVE BEEN PLUGGED INTO THE UPS PROPERLY, WHICH LED TO THE OVERHEATING OF THE ABBOTT POWER CORD (CORD, PWR, 125VAC/10A). THE ANALYZER RETURNED TO NORMAL OPERATION AFTER REPLACEMENT OF THE ABBOTT POWER CORD (CORD, PWR, 125VAC/10A). THE ABBOTT POWER CORD (CORD, PWR, 125VAC/10A) IS SPECIFICALLY USED WITH THE CELL-DYN RUBY ANALYZER. THERE HAVE BEEN NO FURTHER DOCUMENTED OCCURRENCES OF BURNT COMPONENTS AFTER THE ABBOTT POWER CORD (CORD, PWR, 125VAC/10A) WAS REPLACED. THE INSTRUMENT SERVICE HISTORY WAS PERFORMED AND FOUND NO CONTRIBUTING FACTOR ON OR AROUND THE DATE OF THE EVENT. A COMPLAINT SEARCH FOR THE PAST 12 MONTHS ONLY IDENTIFIED THIS SINGLE EVENT ASSOCIATED WITH A BURNT ABBOTT POWER CORD (CORD, PWR, 125VAC/10A). THE COMPLAINT SEARCH DID IDENTIFY ONE ADDITIONAL COMPLAINT, HOWEVER IT WAS DETERMINED TO BE NOT RELATED TO THE CURRENT COMPLAINT ISSUE AS THE CUSTOMER UTILIZED A NON-ABBOTT POWER CABLE ON THE RUBY ANALYZER. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES FOR THE ABBOTT POWER CORD (CORD, PWR, 125VAC/10A). LABELING WAS REVIEWED AND WAS FOUND TO ADDRESS THE CUSTOMER REPORTED ISSUE. BASED ON THE AVAILABLE INFORMATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED.
IT WAS OBSERVED THAT THE POWER CABLE FROM THE CELL-DYN RUBY ANALYZER THAT PLUGS INTO A UPS WAS FOUND BURNED. THE POWER CABLE WAS REPLACED. THERE WAS NO REPORTED INJURY.
IT WAS OBSERVED THAT THE POWER CABLE FROM THE CELL-DYN RUBY ANALYZER THAT PLUGS INTO A UPS WAS FOUND BURNED. THE POWER CABLE WAS REPLACED. THERE WAS NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 726201 | CELL-DYN RUBY SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | ABBOTT LABORATORIES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |