BD PHOENIX PANEL NMIC-306
Report
- Report Number
- 1119779-2024-00721
- Event Type
- Malfunction
- Date Received
- October 3, 2024
- Date of Event
- August 30, 2024
- Report Date
- October 30, 2024
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 00382904436247
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K022129, K023444, K023634, K023858, K024153, K031530, K031699, K031912, K032299, K032567, K032655, K033362, K033560, K041384, K042932, K052269, K060214, K060217, K060257, K060444, K060447, K061327, K061355, K062207, K062944, K063301, K063486, K063573, K063811, K063824, K071623, K132674, K132909, K151320, K181665, K173252, K190905, K163637, K173523, K033458, AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR HIGH MIC ERTAPENEM (ETP) WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER 4199980. THE CUSTOMER DID NOT PROVIDE ISOLATES OR PANEL RETURNS BUT PROVIDED PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE LAB REPORTS SHOW HIGH MIC ETP WITH KLEBSIELLA PNEUMONIAE, KLEBSIELLA VARIICOLA, PROVIDENCIA STUARTII AND ENTEROBACTER CLOACAE COMPLEX WHEN USING THE COMPLAINT BATCH. TO INVESTIGATE, TWO RETENTION PANELS EACH FROM THE COMPLAINT BATCH WERE INOCULATED WITH IN HOUSE ISOLATES K. PNEUMONIAE ENF 11000, K. VARIICOLA ENF 22211, P. STUARTII ENF 9508 AND E. CLOACAE COMPLEX ENF 9907 THEN PLACED IN A PHOENIX M50 FOR ETP MIC RESULTS. IN ADDITION, ONE CONTROL PANEL EACH FROM THE SAME MATERIAL BUT DIFFERENT BATCH WERE INOCULATED WITH IN HOUSE ISOLATES K. PNEUMONIAE ENF 11000, K. VARIICOLA ENF 22211, P. STUARTII ENF 9508 AND E. CLOACAE COMPLEX ENF 9907 THEN PLACED IN A PHOENIX M50 FOR ETP MIC RESULTS. RESULTS OF THE INVESTIGATION RETURNED ALL PANELS WITH THE EXPECTED SUSCEPTIBLE MIC RESULTS FOR ETP. THIS COMPLAINT IS NOT CONFIRMED. THE BATCH HISTORY RECORD WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY.
IT WAS REPORTED WHILE USING THE BD PHOENIX PANEL NMIC-306 THAT A PATIENT ISOLATE (K. VARIICOLA) FROM A URINE SAMPLE HAD A HIGH MIC FOR THE DRUG ERTAPENEM. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 4.
IT WAS REPORTED WHILE USING THE BD PHOENIX PANEL NMIC-306 THAT A PATIENT ISOLATE (K. VARIICOLA) FROM A URINE SAMPLE HAD A HIGH MIC FOR THE DRUG ERTAPENEM. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 4.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1403580 | BD PHOENIX PANEL NMIC-306 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY | LON | BECTON DICKINSON & CO. (SPARKS) | 4199980 | 00382904436247 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |