FDA Adverse Event Other Summary report: N

ALLEN HUG-U-VAC STEEP TREND POSITIONER

MDR report key: 2035822 · Received March 23, 2011

Report

Report Number
1221538-2011-00007
Event Type
Other
Date Received
March 23, 2011
Date of Event
February 23, 2011
Report Date
February 25, 2011
Manufacturer
ALLEN MEDICAL SYSTEMS
Product Code
FWZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE CHARGE NURSE STATED THAT THE STAFF USED SHOULDER POSITIONERS ON THE TABLE BENEATH THE HUG-U-VAC, AS DIRECTED IN THE INSTRUCTIONS FOR USE, IN THE TRENDELENBURG CASE BUT THAT THE SUPPORTS HAD NO DIRECT CONTACT WITH THE PATIENT'S SHOULDER AT ANY TIME. THE HOSPITAL REPORTED USING PINK FOAM PADDING ALONG THE PATIENT SHOULDERS, WHICH IS NOT INDICATED IN THE INSTRUCTIONS FOR USE FOR THIS PRODUCT. IT IS UNCLEAR IF THIS WAS A CONTRIBUTING FACTOR. VISUAL INSPECTION AND FUNCTIONAL TESTING OF THE RETURNED UNIT BY AN ENGINEER REVEALED IT WAS FUNCTIONING ACCORDING TO SPECIFICATION.

Description of Event or Problem · 1

FOLLOWING A TRENDELENBURG POSITIONING CASE IN WHICH THE ALLEN HUG-U-VAC STEEP TREND POSITIONER WAS USED, A CUSTOMER CONTACTED THE ALLEN REP POST-OPERATIVELY TO REPORT THE PATIENT WAS EXPERIENCING BI-LATERAL NUMBNESS AND TINGLING IN HER HANDS AND FINGERS. THE PATIENT WAS RELEASED FROM THE HOSPITAL WITH SOME OF THE NUMBNESS AND TINGLING SENSATION CONDITION PERSISTING. ADDITIONAL TREATMENT AND FOLLOW-UP MAY BE REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALLEN HUG-U-VAC STEEP TREND POSITIONER HUG-U-VAC POSITIONER FWZ ALLEN MEDICAL SYSTEMS A-60001-A1 AP1 11 87125

Patients

Seq Age Sex Outcome Treatment
1 UNK Other