Report Number
3002601200-2024-00484
Event Type
Malfunction
Date Received
October 2, 2024
Date of Event
September 4, 2024
Report Date
October 8, 2024
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830831
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEWLOT#3353177 1-THIS BATCH OF PRODUCTS WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN DECEMBER 2023, AND PACKAGED AT CFS PACKAGE LINE IN DECEMBER 2023. WORK ORDER QUANTITY WAS (B)(4) EA. 2-REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. AMONG THEM, THE PRN TORQUES MEET THE OUTGOING INSPECTION REQUIREMENT. 3-REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 4-THE PRN BATCHES USED IN THIS BATCH OF PRODUCTS ARE 3331293, 3353569, 3113461, 3113465, REVIEW THE RAW MATERIAL INSPECTION RECORDS, NO ABNORMALITIES. 2. NO DEFECTIVE SAMPLES AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. THE RETAINED SAMPLE OF THIS BATCH IS TAKEN FOR 45PSI LEAKAGE TEST AND PRN REMOVAL TORQUE TEST, THE LEAKAGE TEST IS QUALIFIED, THERE IS NO LEAKAGE AT THE PRN, AND THE PRN REMOVAL TORQUE IS WITHIN THE PRODUCT SPECIFICATIONS. PLEASE SEE ATTACHMENT FOR THE TEST REPORTS. 4. IN THE ASSEMBLY PROCESS OF PRN, THERE ARE TORQUE AND ASSEMBLY STROKE MONITORING TO ENSURE THAT THE LUER OF PRN CAN BE ASSEMBLED INTO THE PP CONNECTOR TO A CERTAIN DEPTH AND THE THREAD HAS A GOOD FASTENING EFFECT. IF THE PRN IS NOT IN PLACE, THE EQUIPMENT ALARMS AND REMOVES THE PRODUCT. HOWEVER, ALTHOUGH PRN IS FASTENED TO THE PRODUCT DURING ASSEMBLY, IT MAY COME LOOSE IF IT IS VIBRATED DURING TRANSPORTATION. RECOMMENDATION: TIGHTEN THE PRN BEFORE USE TO PREVENT LEAKAGE. 5. NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS RETAINED SAMPLE, AND NO SIMILAR COMPLAINTS HAVE BEEN RECEIVED FROM OTHER HOSPITALS REGARDING THIS BATCH OF PRODUCTS. AS THE DEFECT STATUS OF THE COMPLAINED SAMPLE CANNOT BE IDENTIFIED, AND THE USAGE OF THE SAMPLE IS UNKNOWN, THE ROOT CAUSE OF LEAKAGE AT THE PRN CANNOT BE DETERMINED. THE PLANT WILL CONTINUE TO TRACK AND TREND FOR SUCH DEFECTS.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMAITON PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC LEAKED AT DIAPHRAGM. WHILE PERFORMING A PUNCTURE FOR A PATIENT'S MEDICATION INFUSION, THIS INDWELLING NEEDLE WAS FOUND TO BE LEAKING MEDICATION AT THE HEPARIN CAP AND WAS UNUSABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1870840 BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 3353177 00382903830831

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown