FDA Adverse Event Malfunction Summary report: N

INSYTE AUTOG BC

MDR report key: 20352970 · Received October 2, 2024

Report

Report Number
1710034-2024-01085
Event Type
Malfunction
Date Received
October 2, 2024
Date of Event
September 11, 2024
Report Date
October 8, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825349
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382534 AND LOT NUMBER 4030761. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOG BC NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RN WAS ATTEMPTING TO INSERT IV. IT WAS INTENDED TO START AN IV AND HAVE THE NEEDLE RETRACT INTO THE SAFETY GLIDE. WHEN THE NEEDLE WAS INSERTED AND RETRACTED, THE NEEDLE DID NOT RETRACT INTO THE SAFETY GLIDE. PT HAS PARKINSON'S AND WAS SHAKING, LUCKILY THE LAB TECH WAS ABLE TO GRAB THE EXPOSED NEEDLE AND PLACE IT IN SHARPS BEFORE ANYONE WAS POKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
34839 INSYTE AUTOG BC PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4030761 00382903825349

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown