Report

Report Number
0002023141-2024-03132
Event Type
Injury
Date Received
October 1, 2024
Report Date
February 23, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

SIMILAR COMPLAINTS FOR BONE LOSS RELATED PROBLEMS HAVE BEEN PREVIOUSLY INVESTIGATED. VISUAL AND DIMENSIONAL EVALUATIONS OF THE PREVIOUSLY RETURNED PRODUCT HAVE NOT IDENTIFIED OR SUGGESTED MANUFACTURING NON-CONFORMANCES. FURTHERMORE, THE PROBABILITY OF MANUFACTURING OR DESIGN DEFECTS THAT MIGHT LEAD TO BONE LOSS ESCAPING THE AVAILABLE DETECTIONS IS REMOTE AND ALMOST NONEXISTENT. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT. ADDITIONALLY, ALL DEVICE HISTORY RECORD REVIEWS VERIFIED THAT EACH IMPLANT WAS STERILIZED PER PROCEDURE FOR EVERY DEVICE. ALL COMPLAINT DATA USED FOR THE SUMMARY INVESTIGATION WAS FOUND TO BE CONFORMING AND DID NOT MEET CAPA/HHE/D/ OR ANY FURTHER ESCALATIONS. THEREFORE, THERE WERE NO COMPLAINTS WHICH CONFIRMED A MANUFACTURING OR DESIGN RELATED ISSUE THAT DID OR COULD CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE DHR WAS REQUESTED, HOWEVER AN ELECTRONIC COPY IS NOT AVAILABLE FOR REVIEW. THE INVESTIGATION WILL BE REOPENED AND UPDATED UPON ANY NEW OR ADDITIONAL INFORMATION PROVIDED TO THIS COMPLAINT. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO ENSURE THE DISTRIBUTION OF CONFORMING PRODUCT IS WITHIN SPECIFICATIONS. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 970302 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: ¿BONE LOSS¿ AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THE REPORTED EVENT LIKELY EXIST OUTSIDE OF ZIMVIE CONTROL, INCLUDING THOSE RELATED TO MEDICAL CONDITIONS / PATIENT HABITS AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION AND NO ASSOCIATION BETWEEN THE DENTAL IMPLANT AND BONE LOSS WAS FOUND THAT CAN EXPLAIN THESE EVENTS. THEREFORE, THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION. MONTHLY TRENDING TO DATE HAS NOT IDENTIFIED ANY STATISTICAL TRIGGERS SUGGESTING POTENTIAL DESIGN OR MANUFACTURING RELATED ISSUES. PREVIOUSLY COMPLETED INVESTIGATIONS FOR BONE LOSS PROBLEMS REPORTED IN ASSOCIATION WITH IMPLANT FAILURE HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES IMPACTING THE MANUFACTURING AND STERILIZATION PROCESSES. THEREFORE, ESCALATION TO CAPA IS NOT REQUIRED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION. THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT G3: DATE RECEIVED BY MANUFACTURER G6: CHECKED "FOLLOW-UP" H3: CHANGED "YES" TO "NO" H6: ENTERED EVALUATION CODES H10: ADDED MANUFACTURER NARRATIVE H3 OTHER TEXT : SUMMARY INVESTIGATION

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. B3: DATE OF EVENT UNKNOWN / NOT PROVIDED . D4: DEVICE EXPIRATION DATE UNKNOWN / NOT PROVIDED. D4: DEVICE UDI NUMBER UNKNOWN / NOT PROVIDED. E1: LAST NAME UNKNOWN / NOT PROVIDED. G4: PMA/510(K) NUMBER UNKNOWN / NOT PROVIDED. H4: DEVICE MANUFACTURER DATE UNKNOWN / NOT PROVIDED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #3 WAS REMOVED DUE TO BONE LOSS. PATIENT RETURNED AFTER SEVERAL YEARS WITH BONE LOSS AND IMPLANT NEEDED TO BE REMOVED SYMPTOMS AS A RESULT OF THE EVENT: INFLAMMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2161950 SPLINE HA CYLINDER 4.0 10 MM DENTAL IMPLANT DZE ZIMMER DENTAL 970302

Patients

Seq Age Sex Outcome Treatment
1 61 YR Female Required Intervention