FDA Adverse Event Malfunction Summary report: N

CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V

MDR report key: 20347570 · Received October 1, 2024

Report

Report Number
2249723-2024-0003996
Event Type
Malfunction
Date Received
October 1, 2024
Date of Event
September 9, 2024
Report Date
February 20, 2025
Manufacturer
DATASCOPE CORP.
Product Code
DSP
UDI-DI
10607567107394
PMA / PMN Number
K063525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION. SUSPECT DEVICE ORIGINALLY DISTRIBUTED AS CS100, UPGRADED TO CS300 USING CONVERSION KIT. UDI AND PRODUCT INFORMATION PROVIDED FOR ORIGINAL MODEL/CATALOG, AS PER PRODUCT LABELING.

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B4, D8, D9, G3, G6, H2, H3, H6 ( TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, COMPONENT CODES, INVESTIGATION CONCLUSION), H11 2249723-2024-0003996. FIELDS D4 FROM INITIAL EMDR ARE FOR ORIGINAL IABP CS100; HOWEVER, THIS IABP WAS UPGRADED TO A CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V CATALOG 0998-00-3023-53. UDI (B)(4), WHICH WAS REPORTED BY THE CUSTOMER. A GETINGE FIELD SERVICE ENGINEER (FSE) ISOLATED FAILURE TO VIDEO RECEIVER PCB. REPLACED VIDEO RECEIVER PCB. AS A PRECAUTIONARY MEASURE, REPLACED DISPLAY CONTROLLER PCB, LUER FTTING, AND DC TO AC INVERTER PCB. PERFORMED COMPLETE PM WITH FULL CALIBRATION. UNIT PASSES ALL FUNCTIONAL AND SAFETY TESTS PER FACTORY SPECIFICATIONS. RETURNED TO CUSTOMER AND CLEARED FOR CLINICAL USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE ON A PATIENT, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAS LINES ON THE DISPLAY. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

N/A

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2369579 CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. 0998-00-3013-53 10607567107394

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown