FDA Adverse Event Other Summary report: N

FLEXIBLE CYSTOSCOPIC FOREIGN BODY FORCEPS

MDR report key: 2034687 · Received March 18, 2011

Report

Report Number
1519132-2011-00011
Event Type
Other
Date Received
March 18, 2011
Date of Event
February 11, 2011
Report Date
March 18, 2011
Manufacturer
GYRUS ACMI, INC.
Product Code
FCL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

VISUAL EXAMINATION OF THE INSTRUMENT FINDS THE DISTAL END OF THE PUSH ROD TO BE BROKEN FROM THE LINK ASSEMBLY. THE LINK ASSEMBLY AND RIVET IS STILL INTACT. THE DISTAL END OF THE PUSH ROD WAS NOT RETURNED WITH THE INSTRUMENT. THE DISTAL END OF THE FLEXIBLE SHAFT ASSEMBLY IS COMPLETELY SEVERED FROM THE JAW ASSEMBLY. THE FLEXIBLE SHAFT ASSEMBLY EXHIBITS SIGNS OF BEING STRETCHED APART AND TWISTED TO THE POINT OF MATERIAL FAILURE. THERE ARE NO OTHER SIGNS OF DAMAGE ON THE BALANCE OF THE INSTRUMENT. THE BROKEN DISTAL END OF THE PUSH ROD IS TYPICALLY AN INDICATION OF EXCESSIVE FORCE APPLIED TO THE DEVICE. THE BROKEN FLEXIBLE SHAFT ASSEMBLY IS ALSO A RESULT OF EXCESSIVE FORCE HAVING BEEN APPLIED. WITHOUT FURTHER INFORMATION FROM THE CUSTOMER, THE ASSESSMENT FINDS THE FAILURE DUE TO CUSTOMER MISUSE.

Description of Event or Problem · 1

DURING A SURGICAL PROCEDURE THE FLEXIBLE CYSTOSCOPIC FOREIGN BODY FORCEPS BROKE IN THE PATIENT, ALL PIECES WERE RETRIEVED FROM THE PATIENT WITHOUT ANY HARM.