FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOG BC

MDR report key: 20344987 · Received October 1, 2024

Report

Report Number
1710034-2024-01079
Event Type
Malfunction
Date Received
October 1, 2024
Date of Event
July 22, 2024
Report Date
September 30, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825127
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E. FACILITY NAME EXCEEDS CHARACTER LIMIT: (B)(6). THIS MDR IS THE RESULT OF AN INVESTIGATION FINDING: THE COMPLAINT THAT THE CATHETER WAS STICKING WHEN ADVANCED WAS CONFIRMED FROM CIRCUMSTANTIAL EVIDENCE THAT WAS OBSERVED ON THE REPRESENTATIVE 24G INSYTE AUTOGUARD DEVICES THAT WERE RECEIVED IN SEALED PACKAGING FROM LOT #4143183. THE AFFECTED UNIT WAS NOT PROVIDED FOR INVESTIGATION. DEFORMATION WAS OBSERVED WITHIN ONE CATHETER ADAPTER, WHICH PUSHED MATERIAL FROM THE OPENING OF THE CATHETER ADAPTER INTO THE LUER TAPER. THE EXTRA MATERIAL APPEARED TO CAUSE RESISTANCE BETWEEN THE CATHETER ADAPTER AND THE NEEDLE HUB WHEN ATTEMPTING TO ADVANCE THE IV CATHETER OFF THE NEEDLE HUB. AS THE SAMPLE WAS RECEIVED IN SEALED PACKAGING, THE DAMAGE LIKELY OCCURRED DURING MANUFACTURING. MISALIGNMENT DURING ASSEMBLY MAY HAVE BEEN A CONTRIBUTING FACTOR. THE APPROPRIATE MANUFACTURING PERSONNEL WERE NOTIFIED OF THIS COMPLAINT AND ADDITIONAL COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. NO OTHER SIMILAR COMPLAINTS WERE REPORTED FROM THIS LOT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOG BC ADAPTER/CONNECTOR IS DEFECTIVE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CATHETER SHEATH ¿STICKS¿ WHEN WE TRY TO ADVANCE IT. I¿VE TRIED SEVERAL NEEDLES AND THE PROBLEM IS ABLE TO BE REPLICATED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40129 BD INSYTE AUTOG BC PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4143183 00382903825127

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown