(B)(4). 3004753838-2024-251916 WAS REPORTED IN ERROR. PLEASE DISREGARD INITIAL REPORTING OF THIS EVENT AS THIS EVENT HAS NOW BEEN DEEMED NOT REPORTABLE. B5: DESCRIBE EVENT OR PROBLEM - SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT. H2 TYPE OF FOLLOW UP- CORRECTION.

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OCCURRED. PRODUCT HAS BEEN RECEIVED BUT IS PENDING EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION. PERFORMANCE DATA WAS REVIEWED. THE ALLEGATION WAS CONFIRMED DUE TO THE FINDING OF SIGNAL LOSS OVER AN HOUR WITHIN THE INVESTIGATION WINDOW. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35420 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 17807570 00386270001788

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female