HOMECHOICE
Report
- Report Number
- 1423500-2011-03836
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Date of Event
- March 7, 2011
- Report Date
- March 8, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER; HOWEVER, THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). EVALUATION SUMMARY: THE DEVICE FAILED THE RITE (RETURN INSTRUMENT TEST/EVALUATION) DUE TO GROUND BOND FAILED PERFORMANCE SPECIFICATION. THE CAUSE WAS DETERMINED TO BE A LOOSE SETSCREW ON THE DOOR POST. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE RITE FAILURE. THE DOOR POST AND DOOR COVER WILL BE SCRAPPED DURING SERVICE. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
DURING EVALUATION OF THE HOMECHOICE (HC) DEVICE, THE PRODUCT ANALYSIS LABORATORY DETERMINED THE HC MACHINE SYSTEM FAILED RETURNED INSTRUMENT TEST/EVALUATION TESTING DUE TO THE FOLLOWING FAILURE: GROUND BOND FAILED PERFORMANCE SPECIFICATION. NO ALLEGATIONS WERE MADE AGAINST ANY OF THE PATIENT'S DIALYSIS PRODUCTS. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED. NO FURTHER INFORMATION WAS PROVIDED.