PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2033973
·
Received March 18, 2011
Report
- Report Number
- 2032227-2011-00714
- Event Type
- Injury
- Date Received
- March 18, 2011
- Date of Event
- March 1, 2011
- Report Date
- March 7, 2011
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT SHE WAS HOSPITALIZED DUE TO A LOW BLOOD GLUCOSE READING OF 40 MG/DL. THE CUSTOMER STATED THAT SHE EXPERIENCED HYPOGLYCEMIA DUE TO NOT EATING. THE CUSTOMER ALSO REPORTED NO DELIVERY ALARMS PRIOR TO THE EVENT. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP PASSED THE PRIME TEST, BUT FAILED THE HIGH PRESSURE TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC MINIMED | MMT-522RNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Hospitalization |