FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2011-02711
- Event Type
- Malfunction
- Date Received
- March 30, 2011
- Date of Event
- March 12, 2011
- Report Date
- May 10, 2011
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THIS IS AN INITIAL REPORT. THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. NOTE: THE METER IS DESIGNED TO REPORT READINGS OF 20 MG/DL TO 500 MG/DL. IT SHOULD BE NOTED THAT THIS METER DOES NOT GIVE NUMERIC VALUE FOR READINGS LESS THAN 20 MG/DL. A "LO" DISPLAY MESSAGE INDICATES A READING LESS THAN 20 MG/DL.
THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE NOT FOUND IN THE METER MEMORY WITHIN 10 MINUTES.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF "LO" (LESS THAN 20 MG/DL) AND 90 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.