ADJ PIN COLLET 2.0-3.2MM
MDR report key: 2033952
·
Received March 23, 2011
Report
- Report Number
- 1811755-2011-00898
- Event Type
- Malfunction
- Date Received
- March 23, 2011
- Date of Event
- February 23, 2011
- Report Date
- February 23, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT RELEASE THE PIN DURING A PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACK UP DEVICE WITH NO ALLEGATION OF ADVERSE CONSEQUENCES.