FDA Adverse Event Malfunction Summary report: N

ADJ PIN COLLET 2.0-3.2MM

MDR report key: 2033952 · Received March 23, 2011

Report

Report Number
1811755-2011-00898
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
February 23, 2011
Report Date
February 23, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT RELEASE THE PIN DURING A PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACK UP DEVICE WITH NO ALLEGATION OF ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADJ PIN COLLET 2.0-3.2MM INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK