FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2033943 · Received March 18, 2011

Report

Report Number
3004209178-2011-80726
Event Type
Injury
Date Received
March 18, 2011
Date of Event
March 4, 2011
Report Date
March 17, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE BUTTON ERROR ALARM AND INTERMITTENT BUTTON RESPONSE WAS DUE TO MOISTURE DAMAGE ON THE BUTTON KEYPAD TRACE. NO BLANK DISPLAY OR ALARM ANOMALIES WERE NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WENT TO THE EMERGENCY ROOM. THE CUSTOMER STATED THAT ONE DAY PRIOR TO THE EVENT, THE INSULIN PUMP WAS NOT WORKING AFTER GIVING A LONG BUTTON ERROR ALARM. THE CUSTOMER FURTHER STATED THAT THE BUTTONS WERE UNRESPONSIVE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 21 YR Hospitalization