FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2033943
·
Received March 18, 2011
Report
- Report Number
- 3004209178-2011-80726
- Event Type
- Injury
- Date Received
- March 18, 2011
- Date of Event
- March 4, 2011
- Report Date
- March 17, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE BUTTON ERROR ALARM AND INTERMITTENT BUTTON RESPONSE WAS DUE TO MOISTURE DAMAGE ON THE BUTTON KEYPAD TRACE. NO BLANK DISPLAY OR ALARM ANOMALIES WERE NOTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WENT TO THE EMERGENCY ROOM. THE CUSTOMER STATED THAT ONE DAY PRIOR TO THE EVENT, THE INSULIN PUMP WAS NOT WORKING AFTER GIVING A LONG BUTTON ERROR ALARM. THE CUSTOMER FURTHER STATED THAT THE BUTTONS WERE UNRESPONSIVE. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR | Hospitalization |