FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2033853 · Received March 30, 2011

Report

Report Number
2024168-2011-02162
Event Type
Injury
Date Received
March 30, 2011
Date of Event
March 3, 2011
Report Date
March 7, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). IT WAS INITIALLY REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS. SUBSEQUENT INFORMATION REVEALED THAT THE DEVICE WAS DISCARDED AND IS NOT AVAILABLE FOR EVALUATION. A CUFF MISS CAN OCCUR DUE TO A NUMBER OF FACTORS INCLUDING, BUT NOT LIMITED TO, NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO INTERACTION WITH HUMAN TISSUE, AGGRESSIVE AND FAST DEPLOYMENT OF THE PLUNGER OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT DURING NEEDLE DEPLOYMENT. THE RETURN OF THE PROGLIDE DEVICE MAY HAVE AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED CUFF MISS. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, A 100% IN PROCESS TESTING OF DEVICES ARE PERFORMED TO ASSURE THERE IS NO NEEDLE DEFLECTION, WHICH MAY CAUSE THE EVENT REPORTED IN THIS CASE. A SAMPLING OF FINISHED DEVICES IS ALSO TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. BASED ON THE INFORMATION RECEIVED WITH THIS INCIDENT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED EXPERIENCE CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 1

(B)(4): DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. ESTIMATED DATE (REPORTED AS OCCURRED SOMETIME LAST WEEK). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICAIN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 970046H

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention