FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2033842 · Received March 23, 2011

Report

Report Number
1824206-2011-01790
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
March 1, 2011
Report Date
March 1, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND A FAULTY CPR VALVE. THE CPR WAS STILL WORKING. HE REPLACED THE CPR VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE HEAD OF THE BED IS DRIFTING SLOWLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1