FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2033829 · Received March 23, 2011

Report

Report Number
2531779-2011-01851
Event Type
Malfunction
Date Received
March 23, 2011
Report Date
February 24, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. ALL KEYPAD BUTTONS REQUIRED EXCESSIVE FORCE TO ACTIVATE A RESPONSE DURING TESTING. DURING INVESTIGATION, THE UP ARROW AND DOWN ARROW KEY CONTACTS WERE OBSERVED TO BE MISALIGNED. ALL KEYPAD BUTTONS INTERMITTENTLY SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP ARROW, DOWN ARROW, OK, AND CONTRAST BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE A RESPONSE. IT WAS REPORTED THAT THE PUMP IS NOT EXPOSED TO WATER, AND THERE ARE NO VISIBLE TEARS OR PEELING ON KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1