FDA Adverse Event Malfunction Summary report: N

4 MOTOR ADVANCE BED

MDR report key: 2033826 · Received March 23, 2011

Report

Report Number
1824206-2011-01777
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
February 28, 2011
Report Date
February 28, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE OUTER CONTROL PANEL ASSEMBLY WAS NOT FUNCTIONING CORRECTLY. HE REPLACED THE OUTER CONTROL ASSEMBLY TO REPAIR THE BED.

Description of Event or Problem · 1

THE TECH FOUND THE BED MOTOR LOCKOUTS WERE ENGAGED. WHEN HE UNLOCKED THE MOTORS, THE FUNCTIONS WOULD NOT STOP (THE HEAD SECTION WOULD RISE UNINTENTIONALLY) UNLESS THE LOCKOUTS WERE ENGAGED AGAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4 MOTOR ADVANCE BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1115

Patients

Seq Age Sex Outcome Treatment
1 UNK