FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 2033818 · Received March 23, 2011

Report

Report Number
1824206-2011-01787
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
February 28, 2011
Report Date
February 28, 2011
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT WILL REPLACE THE CABLE. REPAIRS HAVE NOT BEEN COMPLETED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE BED IS ALARMING A SERVICE CODE 10-019. HE CHANGED THE LEFT UCM CABLE AND THE PROBLEM RETURNED. HE NOW HAS SERVICE CODES 10-19 AND 10-70. THE ACCOUNT FOUND THE SIDECOM CABLE WAS CUT WHICH CAUSED THE SERVICE CODES AND WIRING WAS SHOWING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1