FDA Adverse Event
Malfunction
Summary report: N
MICRO SAGITTAL SAW
MDR report key: 2033744
·
Received March 23, 2011
Report
- Report Number
- 1811755-2011-00923
- Event Type
- Malfunction
- Date Received
- March 23, 2011
- Date of Event
- February 26, 2011
- Report Date
- February 26, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
DURING THE QUALITY INVESTIGATION TESTING, THE CUSTOMER'S CONCERN WAS CONFIRMED; THE DEVICE WAS RUNNING WITHOUT ACTIVATION. FURTHER INVESTIGATION REVEALED A BROKEN BEARING AND CORROSION DAMAGE TO THE MID SPEED MOTOR AND THE PRESS PLUG. THE CORRODED MID SPEED MOTOR WAS IDENTIFIED AS THE PRIMARY CAUSE OF THE FAILURE. THE FAULTY PARTS WERE REPLACED AND THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
A STRYKER MICRO SAGITTAL SAW WAS SENT IN FOR REPAIR BECAUSE, IT WAS RUNNING ON ITS OWN. THE EVENT WAS DISCOVERED DURING ROUTINE MAINTENANCE TESTING. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO SAGITTAL SAW | BONE CUTTING INSTRUMENT AND ACCESSORIES | HBE | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |