FDA Adverse Event Malfunction Summary report: N

COOL PATH DUO 7F, 1304-CP2-7-1.5(5)-L1-TE4BE1AB

MDR report key: 2033639 · Received February 11, 2011

Report

Report Number
2030404-2011-00036
Event Type
Malfunction
Date Received
February 11, 2011
Date of Event
January 17, 2011
Report Date
January 17, 2011
Manufacturer
ST. JUDE MEDICAL, IRVINE
Product Code
OAD
PMA / PMN Number
P060019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS HAVE CONFIRMED A LEAK IN THE HANDLE OF THE CATHETER IS CAUSED BY THE LUMEN BREAKING AT THE EDGE OF THE CATHETER HANDLE IN THE PROTECTING TUBE. REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THIS LOT MET MFG REQUIREMENTS PRIOR TO SHIPMENT. A QUALITY IMPROVEMENT PROJECT WAS INITIATED TO ADDRESS THIS ISSUE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THERE WAS A HANDLE LEAK. THERE WERE NO CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOL PATH DUO 7F, 1304-CP2-7-1.5(5)-L1-TE4BE1AB NONE OAD ST. JUDE MEDICAL, IRVINE 83565 3198427

Patients

Seq Age Sex Outcome Treatment
1 UNK