HEARTSTRING III PROXIMAL SEAL
Report
- Report Number
- 2242352-2011-00046
- Event Type
- Malfunction
- Date Received
- February 11, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 14, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED INSIDE THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE LOADING DEVICE, UNDER THE WINDOW. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DRESSED ON THE DELIVERY DEVICE. THERE WAS SLIGHT EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THESE FINDINGS, THE REPORTED FAILURE "SEAL DID NOT LOAD PROPERLY" WAS CONFIRMED. AS STATED IN THE IFU, THE USER SHOULD ENSURE THAT THE LOCK IS LOCKED. IF THE LOCK IS UNLOCKED, CARE SHOULD BE TAKEN TO AVOID ANY ACCIDENTAL DEPRESSING OF THE PLUNGER. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE SEAL DID NOT LOAD PROPERLY. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTRING III PROXIMAL SEAL | PROXIMAL SEAL | DXC | MAQUET CARDIOVASCULAR, LLC | HS-3045 | 25022382 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |