FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL

MDR report key: 2033635 · Received February 11, 2011

Report

Report Number
2242352-2011-00046
Event Type
Malfunction
Date Received
February 11, 2011
Date of Event
January 11, 2011
Report Date
January 14, 2011
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED INSIDE THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE LOADING DEVICE, UNDER THE WINDOW. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE DRESSED ON THE DELIVERY DEVICE. THERE WAS SLIGHT EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THESE FINDINGS, THE REPORTED FAILURE "SEAL DID NOT LOAD PROPERLY" WAS CONFIRMED. AS STATED IN THE IFU, THE USER SHOULD ENSURE THAT THE LOCK IS LOCKED. IF THE LOCK IS UNLOCKED, CARE SHOULD BE TAKEN TO AVOID ANY ACCIDENTAL DEPRESSING OF THE PLUNGER. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE SEAL DID NOT LOAD PROPERLY. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING III PROXIMAL SEAL PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HS-3045 25022382

Patients

Seq Age Sex Outcome Treatment
1 NA