MEDIA AND DISTRIBUTION CENTER (MDC) PACS
Report
- Report Number
- 2954704-2011-00003
- Event Type
- Malfunction
- Date Received
- February 11, 2011
- Date of Event
- October 22, 2010
- Report Date
- February 9, 2011
- Manufacturer
- PHILIPS HEALTHCARE INFORMATICS, INC.
- Product Code
- LLZ
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE MDC PACS IS A SOFTWARE APPLICATION THAT IS USED FOR RECEIVING, MANAGING, ARCHIVING, DISTRIBUTING AND RECORDING MEDICAL IMAGES ONTO PORTABLE DIGITAL MEDIA (INCLUDING BUT NOT LIMITED TO COMPACT DISK (CD) AND DVD). PRODUCT EVAL DETERMINED THE PRODUCT OPERATED AS DESIGNED. THE REPORTED MALFUNCTION HAS BEEN ASSESSED IN A HEALTH HAZARD EVAL, AND IT WAS DETERMINED THAT THE RISK IS ACCEPTABLE BASED ON ADDITIONAL ANALYSIS THAT SUPPORTS THE ISSUE OCCURRENCE AS A RESULT OF THE ACTIONS OF A SINGLE UNTRAINED INSTALLER, WHO DID NOT COMPLETELY REVIEW INSTALLATION DOCUMENTATION AND DID NOT CONFIGURE THE PRODUCT IN ACCORDANCE WITH THE SITE WORKFLOW. NO OTHER SITES WITH THIS WORKFLOW HAVE BEEN INSTALLED BY THIS UNTRAINED INSTALLER, AND NO OTHER SITES HAVE REPORTED THIS ISSUE. IF THE SYSTEM WAS MISCONFIGURED IN THIS MANNER AT A SITE THAT DOES NOT USE UNIQUE PT IDENTIFIERS, IT IS EXPECTED THAT THE ISSUE WOULD BE DETECTED QUICKLY BY THE USERS.
THE PROBLEM WAS REPORTED BY THE USER SITE DURING CLINICAL USE: THE IMAGES TRANSFERRED TO THE PACS (PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS), AND ALL MR IMAGES/STUDY WERE STORED IN (B)(6) UNDER THE SAME PT NAME. UPON INVESTIGATION IT WAS DISCOVERED THAT MULTIPLE PT RECORDS WERE INCORRECTLY MERGED IN MDC PACS DUE TO A CONFIGURATION THAT DOES NOT MATCH TO THE SITE WORKFLOW. THE REPORTING SITE DOES NOT ISSUE UNIQUE PT IDENTIFIERS IN ITS WORKFLOW, ALTHOUGH THE SYSTEM WAS CONFIGURED TO USE PT ID AS A SINGLE ENTITY ATTRIBUTED FOR IDENTIFICATION OF PT ID. THERE WAS NO REPORT OF PT OR USER HARM ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDIA AND DISTRIBUTION CENTER (MDC) PACS | MDC PACS | LLZ | PHILIPS HEALTHCARE INFORMATICS, INC. | R2X | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |