FDA Adverse Event Injury Summary report: N

SWIFT-LOCK ANCHOR

MDR report key: 2033509 · Received March 24, 2011

Report

Report Number
1627487-2011-00408
Event Type
Injury
Date Received
March 24, 2011
Date of Event
February 24, 2011
Report Date
February 24, 2011
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION DIV.
Product Code
LGW
PMA / PMN Number
K092371
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT # 1627487-2011-00400. THE PT WAS IMPLANTED WITH DUAL PERCUTANEOUS LEADS ON (B)(6) 2010. IT WAS REPORTED THAT SHE WAS UNABLE TO INCREASE THE AMPLITUDE FOR HER STIMULATION VIA THE PROGRAMMER. A DIAGNOSTIC TEST WAS PERFORMED; HOWEVER, NO IMPEDANCE ISSUES WERE OBSERVED. A SUBSEQUENT X-RAY REVEALED A VISUAL ANOMALY WITH THE LEADS NEAR THE ANCHOR SITE. SURGICAL INTERVENTION WILL BE UNDERTAKEN TO RECTIFY THIS MATTER; HOWEVER, A DATE FOR THE PROCEDURE HAS NOT BEEN SET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SWIFT-LOCK ANCHOR SPINAL CORD STIMULATION ANCHOR LGW ST JUDE MEDICAL - NEUROMODULATION DIV. 1192 3081781

Patients

Seq Age Sex Outcome Treatment
1 48 YR Required Intervention IMPLANT:| IPG LEAD ANCHOR: MODEL: 1192