LIFEVEST WCD 4000 SYSTEM
MDR report key: 2033476
·
Received March 23, 2011
Report
- Report Number
- 3002158293-2011-00338
- Event Type
- Malfunction
- Date Received
- March 23, 2011
- Date of Event
- March 3, 2011
- Report Date
- March 17, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BLANK SCREEN) HAS BEEN CONFIRMED. UPON EVALUATION, THE POWER BRICK CONNECTOR WOULD NOT STAY CONNECTED TO THE CHARGER. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY DETERMINED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER BRICK CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
THE PATIENT SERVICE REPRESENTATIVE (PSR) ASSISTING A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT'S BATTERY CHARGER/MODEM WOULD NOT LIGHT UP DESPITE TRYING MULTIPLE POWER OUTLETS. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.