FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2033474 · Received March 23, 2011

Report

Report Number
3002158293-2011-00329
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
February 21, 2011
Report Date
March 16, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 107) HAS BEEN CONFIRMED. UPON EVALUATION, THE TRUNK CABLE CONNECTOR WAS BROKEN. THE ROOT CAUSE FOR THE BROKEN CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE BROKEN CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT'S SPOUSE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE MONITOR SCREEN WAS SAYING TO CALL ZOLL FOR SERVICE. A REVIEW OF THE PATIENT'S DOWNLOAD WAS SHOWING A SERVICE CODE 107. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 77 YR