ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-01876
- Event Type
- Malfunction
- Date Received
- March 23, 2011
- Date of Event
- February 22, 2011
- Report Date
- February 22, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVAL HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.
THE REPORTER, THE PT'S MOTHER, REPORTED THAT ON (B)(6) 2011 THE PT OBTAINED A LOUD VIBRATING ALARM ON THE PUMP; PRIOR TO THIS ALARM THE PUMP SELF-REBOOTED. THE PT OBSERVED THE PUMP REBOOT, AND HAD NO HIGH OR LOW BLOOD GLUCOSE LEVELS. THE PT DID NOT REPORT ANY SYMPTOMS AND DENIED SEEKING MEDICAL ATTENTION. TROUBLESHOOTING REVEALED THERE WERE NO CRACKS OR CORROSION IN THE BATTERY COMPARTMENT, NO DAMAGE TO THE BATTERY CAP, THE THREADS AND O-RING WERE SECURE, AND THE BATTERY CAP WAS TIGHT AND SECURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |