BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC
Report
- Report Number
- 3002601200-2024-00469
- Event Type
- Malfunction
- Date Received
- September 30, 2024
- Date of Event
- September 3, 2024
- Report Date
- October 15, 2024
- Manufacturer
- BD SUZHOU (MDS)
- Product Code
- FOZ
- UDI-DI
- 00382903830831
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
1. THE CUSTOMER RETURNED 1 PHOTO BUT DID NOT RETURN THE DEFECTIVE SAMPLE. THE PHOTO SHOWS BLOOD OOZING FROM THE END OF THE SEPTUM. 2. DHR/BHR REVIEW LOT#4081458 1-THE PRODUCTS OF THIS BATCH WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN APRIL 2024, AND PACKAGED AT R240 PACKAGE LINE AND CFS PACKAGE LINE IN APRIL 2024. WORK ORDER QUANTITY WAS 198,000 EA. 2-REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MET THE PRODUCT SPECIFICATIONS. 3-REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 4-THE PLANT HAS LAUNCHED CAPA TO INVESTIGATE THE LEAKAGE AT THE SEPTUM, 3. AS THE PLANT RECEIVED SIMILAR COMPLAINTS FROM OTHER HOSPITALS ABOUT THIS BATCH OF PRODUCTS, 800MM SIMULATED CLINICAL LEAKAGE TEST WAS CONDUCTED ON THE RETAINED SAMPLES OF THIS BATCH, AND IT WAS FOUND THAT A FEW SAMPLES LEAKED AT THE END OF THE SEPTUM AFTER THE NEEDLE TIP WAS INSERTED INTO THE TEST DEVICE, AND THE LEAKAGE STOPPED AFTER THE NEEDLE CORE WAS PULLED OUT. CONCLUSION(S): NO ABNORMALITY IS FOUND IN THE PRODUCTION PROCESS. THE RETURNED PHOTO SHOWS BLOOD OOZING FROM THE END OF THE SEPTUM. IN RESPONSE TO THIS DEFECT, THE PLANT HAS LAUNCHED CAPA TO TRACE AND INVESTIGATE ITS ROOT CAUSE.
IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
NO ADDITIONAL INFORMATION PROVIDED.
IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC LEAKED AFTER PERFORMING A SUCCESSFUL INDWELLING NEEDLE PUNCTURE, BLOOD RETURN WAS SEEN TO OOZE OUT TOWARDS THE ISOLATION PLUG, WHICH WAS IMMEDIATELY REMOVED AFTERWARDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1041285 | BD INTIMA-II Y 24GAX0.75IN PRN EC SLM NPVC | INTRAVASCULAR CATHETER | FOZ | BD SUZHOU (MDS) | 4081458 | 00382903830831 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |