FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN CARTRIDGE
MDR report key: 2033420
·
Received February 23, 2011
Report
- Report Number
- 2531779-2011-01848
- Event Type
- Malfunction
- Date Received
- February 23, 2011
- Report Date
- February 22, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. THERE IS NO INVESTIGATION NECESSARY. CARTRIDGES WITH LOT# B201576 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.
Description of Event or Problem · 1
THE REPORTER REPORTED THAT THE INSULIN CARTRIDGE WAS LEAKING INTO THE PUMP'S CARTRIDGE COMPARTMENT. THE PT EXPERIENCED NO SYMPTOMS OF HIGH OR LOW BLOOD GLUCOSE LEVELS, HAD NO HIGH OR LOW BLOOD GLUCOSE LEVELS AND DID NOT SEEK MEDICAL ATTENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP CARTRIDGE | LZG | ANIMAS CORP. | IR 1200/1250/2020/OTP | B201576 |