FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2033418 · Received March 23, 2011

Report

Report Number
2531779-2011-01869
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
January 14, 2011
Report Date
February 22, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K032257
Removal / Correction Number
2531779-02/25/11-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CARTRIDGE WAS RETURNED TO ANIMAS. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST WAS PERFORMED AND FLUID WAS OBSERVED LEAKING OUT AT THE PLUNGER END OF THE CARTRIDGE.

Description of Event or Problem · 1

THE REPORTER REPORTED THAT FROM (B)(6), 2011 THE PT OBSERVED AIR BUBBLES IN THE 'BACK END' OF THE INSULIN CARTRIDGE, AND SHE OBTAINED ELEVATED BLOOD GLUCOSE READINGS RANGING FROM 250 MG/DL TO 365 MG/DL. THE PT DID NOT OBSERVE ANY LEAKS FROM THE CARTRIDGE. DURING THIS TIME PERIOD THE PT EXPERIENCED NO SYMPTOMS OF HIGH OR LOW BLOOD GLUCOSE LEVELS, AND SHE DID NOT SEEK ANY MEDICAL ATTENTION. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP CARTRIDGE LZG ANIMAS CORP. IR 1200/1250/2020/OTP B201581

Patients

Seq Age Sex Outcome Treatment
1 63 YR