FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN CARTRIDGE
MDR report key: 2033418
·
Received March 23, 2011
Report
- Report Number
- 2531779-2011-01869
- Event Type
- Malfunction
- Date Received
- March 23, 2011
- Date of Event
- January 14, 2011
- Report Date
- February 22, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CARTRIDGE WAS RETURNED TO ANIMAS. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST WAS PERFORMED AND FLUID WAS OBSERVED LEAKING OUT AT THE PLUNGER END OF THE CARTRIDGE.
Description of Event or Problem · 1
THE REPORTER REPORTED THAT FROM (B)(6), 2011 THE PT OBSERVED AIR BUBBLES IN THE 'BACK END' OF THE INSULIN CARTRIDGE, AND SHE OBTAINED ELEVATED BLOOD GLUCOSE READINGS RANGING FROM 250 MG/DL TO 365 MG/DL. THE PT DID NOT OBSERVE ANY LEAKS FROM THE CARTRIDGE. DURING THIS TIME PERIOD THE PT EXPERIENCED NO SYMPTOMS OF HIGH OR LOW BLOOD GLUCOSE LEVELS, AND SHE DID NOT SEEK ANY MEDICAL ATTENTION. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP CARTRIDGE | LZG | ANIMAS CORP. | IR 1200/1250/2020/OTP | B201581 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |