FDA Adverse Event Malfunction Summary report: N

HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM

MDR report key: 2033389 · Received March 30, 2011

Report

Report Number
3005099803-2011-00958
Event Type
Malfunction
Date Received
March 30, 2011
Date of Event
March 10, 2011
Report Date
March 10, 2011
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
MNB
PMA / PMN Number
P000040
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE HTA PROCEDURE SET WAS RETURNED AND EVALUATED. VISUAL EXAMINATION OF THE DISPOSABLE HEATER CANISTER CONFIRMED THE EVENT. THE CYLINDER WAS CRYSTALIZED 360° AROUND FOR 3/4 OF THE ENTIRE LENGTH. THE PLASTIC THERMISTER HOUSING WAS ALSO CRYSTALIZED. THE ROOT CAUSE CLASSIFICATION FOR THIS EVENT IS UNDETERMINABLE, BASED ON THE RETURNED CONDITION OF THE PRODUCT.

Additional Manufacturer Narrative · 1

THE COMPLAINANT WAS UNABLE TO PROVIDE THE LOT NUMBER. THEREFORE, EXPIRATION AND MANUFACTURING DATES CANNOT BE DETERMINED. ALTHOUGH THE PRODUCT HAS BEEN RETURNED, THE DEVICE HAS NOT YET BEEN EVALUATED. AT THIS TIME, WE ARE UNABLE TO DETERMINE THE RELATIONSHIP BETWEEN THE DEVICE AND THE CAUSE OF THIS EVENT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

A HYDROTHERMABLATION PROCERVA PROCEDURE SET WAS USED DURING A HYDROTHERMABLATION (HTA) PROCEDURE PERFORMED ON (B)(6), 2011. (PATIENT ID, AGE, DATE OF BIRTH AND WEIGHT ARE UNKNOWN).ACCORDING TO THE COMPLAINANT, DURING PREPARATION, THE HEATER CANISTER OVERHEATED AND MELTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED WITH THIS EVENT AND THE CURRENT CONDITION OF THE PATIENT IS REPORTED TO BE "FINE."

Description of Event or Problem · 1

A HYDROTHERMABLATION PROCERVA PROCEDURE SET WAS USED DURING A HYDROTHERMABLATION (HTA) PROCEDURE PERFORMED ON (B)(6), 2011. (PATIENT ID, AGE, DATE OF BIRTH AND WEIGHT ARE UNKNOWN). ACCORDING TO THE COMPLAINANT, DURING PREPARATION, THE HEATER CANISTER OVERHEATED AND MELTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED WITH THIS EVENT AND THE CURRENT CONDITION OF THE PATIENT IS REPORTED TO BE "FINE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM DEVICE, THERMAL ABLATION, ENDOMETRIAL MNB BOSTON SCIENTIFIC - MARLBOROUGH M006560210

Patients

Seq Age Sex Outcome Treatment
1