FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2033335 · Received March 21, 2011

Report

Report Number
2531779-2011-01814
Event Type
Malfunction
Date Received
March 21, 2011
Date of Event
February 22, 2011
Report Date
February 24, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION REVEALED THAT THERE WAS CONTAMINATION UNDER THE DOWN BUTTON CONTACT. THE DOWN BUTTON INTERMITTENTLY ACTIVATED PUMP FUNCTIONS WHEN PRESSED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE DISTRIBUTOR REPORTED THAT THE DOWN ARROW BUTTON WAS STICKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1