CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM
Report
- Report Number
- 1030489-2024-01213
- Event Type
- Malfunction
- Date Received
- September 30, 2024
- Date of Event
- September 10, 2024
- Report Date
- November 8, 2024
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- MAX
- UDI-DI
- 00763000253554
- PMA / PMN Number
- K210425
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
H3: PRODUCT ANALYSIS: PART# 6061002 LOT# K22K1136 VISUAL AND OPTICAL EXAMINATION IDENTIFIED IT APPEARS THAT PART OF THE TIP OF THE INSTRUMENT HAS BEEN BROKEN OFF. THE METAL DEFORMATION GIVES INDICATION THAT THE FAILURE WAS THE RESULT OF OVERLOAD. H6: FDM AND FDR CODES UPDATED MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED PRIOR TO USE. IT WAS REPORTED THAT THE TIP OF THE DRIVER WAS BROKEN OFF SO WHEN THE SURGEON WENT TO EXPAND THE CAGE, THE DRIVER WAS JUST SPINNING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1053949 | CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM | INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, | MAX | MEDTRONIC SOFAMOR DANEK USA, INC | 6061002 | K22K1136 | 00763000253554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |