FDA Adverse Event Malfunction Summary report: N

CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM

MDR report key: 20332335 · Received September 30, 2024

Report

Report Number
1030489-2024-01213
Event Type
Malfunction
Date Received
September 30, 2024
Date of Event
September 10, 2024
Report Date
November 8, 2024
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
MAX
UDI-DI
00763000253554
PMA / PMN Number
K210425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS: PART# 6061002 LOT# K22K1136 VISUAL AND OPTICAL EXAMINATION IDENTIFIED IT APPEARS THAT PART OF THE TIP OF THE INSTRUMENT HAS BEEN BROKEN OFF. THE METAL DEFORMATION GIVES INDICATION THAT THE FAILURE WAS THE RESULT OF OVERLOAD. H6: FDM AND FDR CODES UPDATED MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED PRIOR TO USE. IT WAS REPORTED THAT THE TIP OF THE DRIVER WAS BROKEN OFF SO WHEN THE SURGEON WENT TO EXPAND THE CAGE, THE DRIVER WAS JUST SPINNING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1053949 CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, MAX MEDTRONIC SOFAMOR DANEK USA, INC 6061002 K22K1136 00763000253554

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown