FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2033177
·
Received February 28, 2011
Report
- Report Number
- 2531779-2011-01190
- Event Type
- Malfunction
- Date Received
- February 28, 2011
- Report Date
- January 29, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE UP/DOWN/OK KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING, THE CONTRAST KEYPAD BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE UP/CONTRAST KEY CONTACTS WERE MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED, THE UP/DOWN/OK KEY CONTACTS BECAME INVERTED DURING TESTING AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE UP ARROW/OK KEYPAD BUTTONS HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |