FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2033083
·
Received February 28, 2011
Report
- Report Number
- 3002158293-2011-00226
- Event Type
- Malfunction
- Date Received
- February 28, 2011
- Date of Event
- January 3, 2011
- Report Date
- February 24, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BROKEN TRUNK CONNECTOR) HAS BEEN CONFIRMED. UPON EVAL, THE TRUNK CABLE CONNECTOR WAS DAMAGED. THERE WERE NO CUT WIRES. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE BROKEN CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
DURING SERVICING OF ELECTRODE BELT SN (B)(4) FOR AN UNRELATED COMPLAINT, A REPORTABLE PROBLEM WAS DISCOVERED. THE BELT WAS FOUND TO HAVE A BROKEN TRUNK CABLE CONNECTOR. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |