FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 2033079 · Received February 28, 2011

Report

Report Number
3004464228-2011-00091
Event Type
Malfunction
Date Received
February 28, 2011
Date of Event
January 29, 2011
Report Date
January 29, 2011
Manufacturer
INSULET CORP
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION OF THE RETURNED POD FOUND EVIDENCE OF AN INTERNAL FLUID LEAK. DISCOLORATION WAS SEEN INSIDE THE DEVICE, AND RESIDUAL FLUID AND CORROSION WERE FOUND ON THE SURFACES OF INTERNAL ASSEMBLIES. A LEAK TEST WAS PERFORMED, WHICH CONFIRMED THE PRESENCE OF A LEAK IN THE FLUID PATH CAUSED BY A TEAR IN THE CANNULA TUBING. THE LEAK WOULD HAVE RESULTED IN INSULIN COMING INTO CONTACT WITH INTERNAL COMPONENTS AND ASSEMBLIES, ULTIMATELY CAUSING THE DEVICE TO FAIL. A POD MALFUNCTION, THEREFORE, IS CONFIRMED TO HAVE BEEN A CONTRIBUTING FACTOR TO THE CUSTOMER'S HIGH BG LEVELS.

Description of Event or Problem · 1

THE REPORT INDICATED THAT THE CUSTOMER EXPERIENCED CONSISTENTLY HIGH BG LEVELS (286-468MG/DL) WITHIN THE FIRST DAY OF WEARING THE POD. THE CUSTOMER'S HUSBAND OPENED THE POD AND NOTICED THAT "THE BATTERIES LOOKED CORRODED." THE POD WILL BE RETURNED FOR EVAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORP 11200 L30452

Patients

Seq Age Sex Outcome Treatment
1 34 YR Other