OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2011-00091
- Event Type
- Malfunction
- Date Received
- February 28, 2011
- Date of Event
- January 29, 2011
- Report Date
- January 29, 2011
- Manufacturer
- INSULET CORP
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE INVESTIGATION OF THE RETURNED POD FOUND EVIDENCE OF AN INTERNAL FLUID LEAK. DISCOLORATION WAS SEEN INSIDE THE DEVICE, AND RESIDUAL FLUID AND CORROSION WERE FOUND ON THE SURFACES OF INTERNAL ASSEMBLIES. A LEAK TEST WAS PERFORMED, WHICH CONFIRMED THE PRESENCE OF A LEAK IN THE FLUID PATH CAUSED BY A TEAR IN THE CANNULA TUBING. THE LEAK WOULD HAVE RESULTED IN INSULIN COMING INTO CONTACT WITH INTERNAL COMPONENTS AND ASSEMBLIES, ULTIMATELY CAUSING THE DEVICE TO FAIL. A POD MALFUNCTION, THEREFORE, IS CONFIRMED TO HAVE BEEN A CONTRIBUTING FACTOR TO THE CUSTOMER'S HIGH BG LEVELS.
THE REPORT INDICATED THAT THE CUSTOMER EXPERIENCED CONSISTENTLY HIGH BG LEVELS (286-468MG/DL) WITHIN THE FIRST DAY OF WEARING THE POD. THE CUSTOMER'S HUSBAND OPENED THE POD AND NOTICED THAT "THE BATTERIES LOOKED CORRODED." THE POD WILL BE RETURNED FOR EVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORP | 11200 | L30452 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Other |