FDA Adverse Event Malfunction Summary report: N

AUTOFEED HUMIDIFICATION CHAMBER

MDR report key: 2033046 · Received March 29, 2011

Report

Report Number
9611451-2011-00205
Event Type
Malfunction
Date Received
March 29, 2011
Report Date
March 1, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RETURNED COMPLAINT CHAMBER WAS VISUALLY INSPECTED. RESULTS: THE VISUAL INSPECTION REVEALED THAT THERE WAS A BREAK IN BETWEEN THE FEEDSET TUBE AND WHERE IT IS INSERTED INTO THE VENT SPIKE. THE SURFACE OF THE BREAK IS ROUGH. A LOT CHECK REVEALED (B)(4) OTHER SIMILAR COMPLAINT OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: THE SURFACE OF THE BREAK IS ROUGH, SUGGESTING THAT THE TUBING WAS BROKEN BY STRETCHING, PLACED UNDER TENSION OR PULLING RATHER THAN CUTTING. ALL MR290 CHAMBERS ARE PRESSURE TESTED FOR LEAKS BEFORE THEY LEAVE THE PRODUCTION LINE. THIS SUGGESTS THAT THE FEED SET TUBING WAS SPLIT POST-PRODUCTION. IT IS UNCLEAR WHETHER THE TUBING MAY HAVE SUSTAINED DAMAGE PRIOR TO FORCE BEING APPLIED TO THE TUBING. ANY PREVIOUSLY SUSTAINED DAMAGE WOULD HAVE MADE THE TUBING MORE SUSCEPTIBLE TO BREAKING IF THE TUBING WAS PULLED. THE MR290 USER INSTRUCTIONS STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT". OUR MONITORING AND TRENDING OF LEAKING MR290 FEED SET TUBING HAS A RATE OF OCCURRENCE OF (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS WATER LEAKAGE FROM THE WATER FEEDSET OF THE MR290V AUTOFEED HUMIDIFICATION CHAMBERS. THIS WAS FOUND PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS WATER LEAKAGE FROM THE WATER FEEDSET OF THE MR290V AUTOFEED HUMIDIFICATION CHAMBERS. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOFEED HUMIDIFICATION CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 100216

Patients

Seq Age Sex Outcome Treatment
1