AUTOFEED HUMIDIFICATION CHAMBER
Report
- Report Number
- 9611451-2011-00205
- Event Type
- Malfunction
- Date Received
- March 29, 2011
- Report Date
- March 1, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE RETURNED COMPLAINT CHAMBER WAS VISUALLY INSPECTED. RESULTS: THE VISUAL INSPECTION REVEALED THAT THERE WAS A BREAK IN BETWEEN THE FEEDSET TUBE AND WHERE IT IS INSERTED INTO THE VENT SPIKE. THE SURFACE OF THE BREAK IS ROUGH. A LOT CHECK REVEALED (B)(4) OTHER SIMILAR COMPLAINT OF THIS NATURE FOR THIS LOT NUMBER. CONCLUSION: THE SURFACE OF THE BREAK IS ROUGH, SUGGESTING THAT THE TUBING WAS BROKEN BY STRETCHING, PLACED UNDER TENSION OR PULLING RATHER THAN CUTTING. ALL MR290 CHAMBERS ARE PRESSURE TESTED FOR LEAKS BEFORE THEY LEAVE THE PRODUCTION LINE. THIS SUGGESTS THAT THE FEED SET TUBING WAS SPLIT POST-PRODUCTION. IT IS UNCLEAR WHETHER THE TUBING MAY HAVE SUSTAINED DAMAGE PRIOR TO FORCE BEING APPLIED TO THE TUBING. ANY PREVIOUSLY SUSTAINED DAMAGE WOULD HAVE MADE THE TUBING MORE SUSCEPTIBLE TO BREAKING IF THE TUBING WAS PULLED. THE MR290 USER INSTRUCTIONS STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT". OUR MONITORING AND TRENDING OF LEAKING MR290 FEED SET TUBING HAS A RATE OF OCCURRENCE OF (B)(4).
A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS WATER LEAKAGE FROM THE WATER FEEDSET OF THE MR290V AUTOFEED HUMIDIFICATION CHAMBERS. THIS WAS FOUND PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS WATER LEAKAGE FROM THE WATER FEEDSET OF THE MR290V AUTOFEED HUMIDIFICATION CHAMBERS. THIS WAS FOUND PRIOR TO PATIENT USE.