PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2011-02131
- Event Type
- Injury
- Date Received
- March 29, 2011
- Date of Event
- March 1, 2011
- Report Date
- March 4, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. A CUFF MISS (NO SUTURE RETRIEVED/ATTACHED) CAN OCCUR DUE TO A NUMBER OF FACTORS INCLUDING, BUT NOT LIMITED TO NEEDLE DEFLECTION DURING PLUNGER DEPLOYMENT DUE TO INTERACTION WITH HUMAN TISSUE, AGGRESSIVE AND FAST DEPLOYMENT OF THE PLUNGER OR A FAILURE TO MAINTAIN A STABLE POSITION OF THE DEVICE WITH RESPECT TO THE TISSUE TRACT DURING NEEDLE DEPLOYMENT. ULTIMATELY, THE RETURN OF PROGLIDE DEVICE MAY HAVE AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED CUFF MISS. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, A 100% IN PROCESS TESTING OF DEVICES ARE PERFORMED TO ASSURE THERE IS NO NEEDLE DEFLECTION, WHICH MAY CAUSE A CUFF-MISS. A SAMPLING OF FINISHED DEVICES IS ALSO TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. THE REPORTED USE OF THE PROGLIDE DEVICE IN A CALCIFIED ARTERY IS INCONSISTENT WITH THE INSTRUCTIONS FOR USE (IFU). THE IFU, UNDER SPECIAL PATIENT POPULATION STATES THAT THE SAFETY AND EFFECTIVENESS OF THE PERCLOSE PROGLIDE SMC DEVICES HAVE NOT BEEN ESTABLISHED IN PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT ACCESS SITE. IT SHOULD BE NOTED THAT THE PATIENTS HISTORY OF PERIPHERAL VASCULAR DISEASE MAY HAVE ALSO PLAYED A ROLE IN THE REPORTED EXPERIENCE. THE LOT HISTORY RECORD FOR THIS PRODUCT WAS NOT REVIEWED AND A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS NOT PERFORMED BECAUSE THE LOT NUMBER WAS NOT REPORTED AND THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. IN THIS INSTANCE, BASED ON THE INFORMATION RECEIVED WITH THIS COMPLAINT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED EXPERIENCE CANNOT BE DETERMINED.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PROGLIDE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF A MODERATELY CALCIFIED COMMON FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, WHEN THE PLUNGER WAS PULLED BACK NO SUTURES WERE ATTACHED TO THE NEEDLES. MANUAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Required Intervention |