CLEARLINK CONTINU-FLO 3 PORTMANIFOLD 10 DPM NV
Report
- Report Number
- 6000001-2011-02325
- Event Type
- Malfunction
- Date Received
- March 29, 2011
- Date of Event
- October 1, 2010
- Report Date
- October 5, 2010
- Manufacturer
- BAXTER HEALTHCARE - AIBONITO
- Product Code
- FMG
- PMA / PMN Number
- K932512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
(B)(4). A CUSTOMER DID NOT SEND IN AN ACTUAL NOR COMPANION SAMPLE FOR EVALUATION; HOWEVER A SIMILAR PRODUCT WAS SENT IN FOR EVALUATION. DURING THE VISUAL INSPECTION THE REPORTED CONDITION WAS NOT CONFIRMED. A FUNCTIONAL TEST AND PRESSURE TEST AT 8PSI WERE PERFORMED WHERE THE RESULTS WERE SATISFACTORY. THE UNIT WAS EVALUATED FOR HAND PULL TEST WITH SUCCESSFUL RESULTS. THE CAUSE FOR THE UNCONFIRMED CONDITION IS UNKNOWN. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SAMPLE REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. A BATCH REVIEW CANNOT BE PERFORMED SINCE THERE WAS NO LOT NUMBER PROVIDED.
A CUSTOMER REPORTED TO BAXTER PRODUCT SURVEILLANCE OF A SEPARATION ISSUE WITH A CONTINU-FLO SOLUTION SET IN WHICH THERE WAS DIFFICULTY SECURING THE MANIFOLD PORTION OF THE TUBING WHICH CAUSED LEAKAGES IN THE OPERATING ROOM. THE CUSTOMER WAS CONNECTING THIS SET TO A MICRO CLAVE LUER. THE FACILITY IS USING A FIRM PUSH AND TWIST METHOD ALWAYS IN ORDER TO CONNECT. THIS INCIDENT OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION WAS AVAILABLE. THIS IS REPORT 2 OF 5 OF THE SAME REPORTED PROBLEM FROM THIS FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CLEARLINK CONTINU-FLO 3 PORTMANIFOLD 10 DPM NV | STOPCOCK, I.V. SET | FMG | BAXTER HEALTHCARE - AIBONITO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | MICROCLAVE LUER, UNKNOWN SOLUTION |