FDA Adverse Event Injury Summary report: N

CORNEAT EVERPATCH

MDR report key: 20328704 · Received September 27, 2024

Report

Report Number
3029772936-2024-00005
Event Type
Injury
Date Received
September 27, 2024
Date of Event
August 28, 2024
Report Date
September 27, 2024
Manufacturer
CORNEAT VISION, LTD.
Product Code
QWU
UDI-DI
G16010362950
PMA / PMN Number
K223074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORDS WERE REVIEWED FOR THIS MANUFACTURING LOT AND THERE WERE NO DISCREPANCIES OR UNUSUAL FINDINGS THAT RELATE TO THE REPORTED EVENT. A CAPA INVESTIGATION HAS BEEN OPENED TO UNDERSTAND THE EVERPATCH RATE OF WOUND DEHISCENCE, CONJUNCTIVAL EROSION, EVERPATCH EXPOSURE, AND SURGICAL INTERVENTION AND TO UNDERSTAND HOW THIS CORRELATES WITH ALTERNATIVE TREATMENTS FOR SCLERAL REINFORCEMENT TO COVER GLAUCOMA DRAINAGE TUBES (E.G., DONOR TISSUE, PERICARDIUM, OR SCLERAL TUNNEL). THE MAJORITY OF EVERPATCH CUSTOMERS WERE CONTACTED AND ASKED TO SHARE THEIR POSTOPERATIVE CLINICAL EXPERIENCE. BASED ON PRELIMINARY CAPA INVESTIGATION RESULTS, CORNEAT ESTIMATES THE LIKELIHOOD OF OCCURRENCE FOR WOUND DEHISCENCE IS 10-15%, WITH ALL KNOWN CASES BEING SEIDEL NEGATIVE (I.E., NO WOUND LEAK). HOWEVER, SURGEONS OPTED TO PERFORM REVISION SURGERY IN APPROXIMATELY 5-10% OF IMPLANTED PATIENTS. EARLY FINDINGS ALSO SHOW CONJUNCTIVAL EROSIONS HEALED SPONTANEOUSLY (I.E., WITHOUT SURGICAL INTERVENTION) IN APPROXIMATELY 1-5% OF IMPLANTED PATIENTS. IMPORTANTLY, THERE HAVE BEEN NO CONFIRMED REPORTS OF GLAUCOMA DEVICE/TUBE EXPOSURE. THESE CURRENT ADVERSE EVENT RATES FOR THE EVERPATCH COMPARE FAVORABLY TO PUBLISHED RATES FOR WOUND DEHISCENCE USING TISSUE TO COVER THE GLAUCOMA TUBE. GEFFEN ET AL REPORT: "CONJUNCTIVAL DEHISCENCE IS USUALLY A BENIGN, COMMON COMPLICATION AFTER AGV [AHMED GLAUCOMA VALVE] INSERTION. IT DOES NOT NEED REPAIR AS LONG AS THE TUBE IS WELL COVERED. AGV TUBE OR PLATE EXPOSURES ARE LESS COMMON, OCCUR LATER AND WERE PROMPTLY REPAIRED AS PER CURRENT PRACTICE." THE AUTHORS REVIEWED THE MEDICAL CHARTS OF 158 SUBJECTS AND REPORT 33.5% OF SUBJECTS PRESENTED WITH WOUND DEHISCENCE AND 8.9% OF SUBJECTS PRESENTED WITH GLAUCOMA DEVICE EXPOSURES. THERE WERE NO CONJUNCTIVAL COMPLICATIONS IN 57.6% OF SUBJECTS. REFERENCE: GEFFEN N, BUYS YM, SMITH M, ANRAKU A, ALASBALI T, RACHMIEL R ET AL. CONJUNCTIVAL COMPLICATIONS RELATED TO AHMED GLAUCOMA VALVE INSERTION. J GLAUCOMA. 2014;23(2):109-14. CONJUNCTIVAL DEHISCENCE IS A COMMON POSTOPERATIVE ADVERSE EVENT AND AN INHERENT RISK OF GLAUCOMA DRAINAGE DEVICE IMPLANTATION. THE DEVICE INSTRUCTIONS FOR USE IDENTIFIES THE FOLLOWING POSSIBLE COMPLICATION: DETACHMENT OF THE DEVICE FROM SURROUNDING TISSUE IN CASE OF TRAUMA OR WHEN CHRONIC INFLAMMATION IS NOT ADDRESSED. MANUFACTURER REFERENCE#: (B)(4).

Description of Event or Problem · 0

ON AUGUST 28, 2024, THE MANUFACTURER BECAME AWARE OF 2 SERIOUS INJURIES INVOLVING THE EVERPATCH DEVICES WHICH WERE IMPLANTED IN CONJUNCTION WITH GLAUCOMA TUBE SHUNTS. SPECIFICALLY, THE PATIENTS PRESENTED WITH WOUND DEHISCENCE, CONJUNCTIVAL RETRACTION, AND EXPOSURE OF THE EVERPATCH. BOTH PATIENTS UNDERWENT SURGICAL REVISION; ONE EVERPATCH WAS EXPLANTED AND REPLACED BY SCLERA AND ONE EVERPATCH WAS CORRECTED BY ADVANCING THE CONJUNCTIVA AND RE-SUTURING. BOTH PATIENTS ARE DOING WELL POSTOPERATIVELY. ADDITIONAL INFORMATION WAS REQUESTED ON SEPTEMBER 16, 2024.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
963149 CORNEAT EVERPATCH PROSTHESIS, EYELID SPACER/GRAFT, POLYMER QWU CORNEAT VISION, LTD. 1007691 20230713 G16010362950

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention