FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 20328559 · Received September 27, 2024

Report

Report Number
2955842-2024-19844
Event Type
Malfunction
Date Received
September 27, 2024
Date of Event
September 4, 2024
Report Date
September 4, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE FSE REPLACED THE INTEGRATED ELECTRO SURGICAL UNIT (IESU) TO RESOLVE THE REPORTED ISSUE. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE THE IESU ASSOCIATED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS (FA). THE IESU WAS ANALYZED, BUT THE REPORTED ISSUE COULD NOT BE REPRODUCED. THE UNIT ENERGIZED AND CAUTERIZED AND ALL PORTS RECOGNIZED INSTRUMENTS. MULTIPLE C-00 ERRORS WERE FOUND IN ERROR LOGS.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE ERBE GENERATOR HAD ERRORS. INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) CONFIRMED 25913 (C-01) ERRORS ON THE ERBE GENERATOR. THE TSE WALKED THE CUSTOMER THROUGH POWER CYCLING THE GENERATOR AND DISCONNECTING THE GENERATOR'S EXTERNAL FOOT PEDALS. THE GENERATOR CAME BACK UP WITH A U-02 ERROR. THE TSE INFORMED THE CUSTOMER THAT THE LOCAL FIELD SERVICE ENGINEER WOULD NEED TO BE DISPATCHED FOR ADDITIONAL TROUBLESHOOTING. THE CUSTOMER PLACED THE CALL ON HOLD AS THEY LOCATED A THIRD-PARTY ESU DEVICE TO CONNECT BUT THE CALL WAS DROPPED. INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE CUSTOMER AND GATHERED THE FOLLOWING INFORMATION: THE SURGEON DECIDED TO CONVERT TO LAPAROSCOPIC SURGERY BECAUSE THE ERBE GENERATOR WAS NOT WORKING. THE ERBE GENERATOR WAS NOT DELIVERING ENERGY. THE PROCEDURE WAS COMPLETED WITH NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2603179 DAVINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-36 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DA VINCI INSTRUMENTS AND ACCESSORIES