STERLING BALLOON DILATATION CATHETER
Report
- Report Number
- 2134265-2011-01123
- Event Type
- Injury
- Date Received
- March 29, 2011
- Date of Event
- February 25, 2011
- Report Date
- March 8, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- DQY
- PMA / PMN Number
- K053118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: (B)(6). DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE ROOT CAUSE IS ANTICIPATED PROCEDURAL COMPLICATIONS AS THIS EVENT IS A KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AND IS NOTED WITHIN THE DFU. (B)(4).
SAME CASE AS MFR REPORT #: 2134265-2011-00869, 2134265-2011-00870, 2134265-2011-01122. IT WAS REPORTED THAT DURING A CAROTID STENTING PROCEDURE, THE PATIENT EXPERIENCED SLOW FLOW AND A STROKE. THE PROCEDURE TREATED THE 80% STENOSED, LONG TARGET LESION LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED COMMON CAROTID ARTERY TO THE INTERNAL CAROTID ARTERY. TREATMENT UTILIZED A FILTERWIRE EZ AND CONSISTED OF PRE-DIALTION WITH A 3.0X30MM STERLING BALLOON, THE PLACEMENT OF A 10X31MM CAROTID WALLSTENT AND POST DILATION WITH A 4.5X30MM STERLING BALLOON. IT WAS NOTED THAT THE VESSEL WAS "OPENED AS INTENDED" AFTER THE PROCEDURE. AFTER PLACING THE CAROTID WALLSTENT, ANGIOGRAPHY REVEALED SLOW FLOW DUE TO A PLAQUE SHIFT. THE PATIENT EXPERIENCED PROGRESSIVE YAWNING AND PALSY AFTER THE PROCEDURE AND WAS GIVEN MEDICATION. FOUR DAYS POST THE PROCEDURE, THE PATIENT'S CONDITION WORSENED AND IT WAS REPORTED THAT THE PATIENT HAD AN EXTENSIVE STROKE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STERLING BALLOON DILATATION CATHETER | CATHETER, PERCUTANEOUS | DQY | BOSTON SCIENTIFIC - MAPLE GROVE | H74939031453010 | 13748897 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | 3.0X30MM STERLING BALLOON| CAROTID WALLSTENT 10X31| FILTERWIRE EZ |