SYNCHRON LX® I 725 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-00900
- Event Type
- Malfunction
- Date Received
- March 29, 2011
- Date of Event
- March 1, 2011
- Report Date
- March 1, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K023049
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER NOTED SAMPLE REFERENCE AND SAMPLE NOISE ERRORS BCI CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER WITH TROUBLESHOOTING AND RECOMMENDED THE USE OF PERSONAL PROTECTIVE EQUIPMENT BEFORE TROUBLESHOOTING. THE CTS HAD CUSTOMER CLEAN THE EIC (ELECTROLYTE INJECTION CUP) AND CALIBRATE ISE (ION SELECTIVE ELECTRODE). THE CUSTOMER STATED THAT SODIUM (NA) WAS GETTING REFERENCE DRIFTS, AND THE CTS SUGGESTED CUSTOMER TO REMOVE NA REFERENCE ELECTRODE, DRY, TWIRL, DRY PORT, REINSERT, PRIME AND RECALIBRATE. THIS DID NOT RESOLVE THE ISSUE. THE CTS SUGGESTED CUSTOMER TO CHECK THE REAGENT, WHICH THE CUSTOMER FOUND LOW. THE CUSTOMER REPLACED THE ISE REFERENCE AND BUFFER, PRIMED AND RECALIBRATED SUCCESSFULLY. THE CUSTOMER WAS TO CALL IF PROBLEM CONTINUES. NO FURTHER COMPLAINT WAS FILED AS OF (B)(6) 2011. SERVICE WAS NOT DISPATCHED FOR THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO LIQUID SPLATTERS ON SYNCHRON LXI 725 CLINICAL SYSTEM. THE CUSTOMER STATED THAT THE LINE 26 WAS POPPING OFF AND THE PROBE GOES INTO EIC SPLATTERED LIQUID. THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON LX® I 725 CLINICAL SYSTEM | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | LXI 725 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |