FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* TROCAR

MDR report key: 2032810 · Received March 29, 2011

Report

Report Number
3005075853-2011-01250
Event Type
Malfunction
Date Received
March 29, 2011
Date of Event
March 3, 2011
Report Date
March 3, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE SLEEVE WAS RETURNED IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS FUNCTIONALLY TESTED TO DETECT ANY LEAKING ISSUES. UPON EVALUATION OF THE DEVICE, IT WAS FUNCTIONALLY LEAK TESTED AND PASSED. THE DEVICE WAS FULLY FUNCTIONAL ACCORDING TO THE MANUFACTURING REQUIREMENTS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. AS THE OBTURATOR WAS NOT RECEIVED AND IT IS THE PART THAT HAS THE BATCH STAMPED, WE DID NOT RECEIVE A BATCH OR LOT NUMBER FOR THE PRODUCT INVOLVED IN THIS COMPLAINT. THEREFORE, WE WERE UNABLE TO CHECK MANUFACTURING RECORDS FOR ANY RELATED NON-CONFORMANCES.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLON PROCEDURE, THE TROCAR WAS LEAKING. THEY KEPT LOSING PNEUMO. A NEW TROCAR WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1